LEDETECT 96 is a computer controlled microplate reader for 96 well plates. Easy and safe operation take away the hassle from your daily microplate reading jobs. It is based on modern LED technology - no need to worry about lamp replacements anymore.
LED Technology: Instead of lamps and filters, LEDETECT 96 is fitted with up to 6 wavelength-specific LEDplugins. Each plugin contains it's own digital ID, LED-light source, filter and lenses in one easily exchangeable component.
Low power consumption: LEDETECT 96 has a maximum consumption of 12W during reading and a standby consumption of not more than 2W.
Unrivaled optical performance: Using LEDs as light source moves all those known problems with halogen lamps used in other readers to history. Modern LEDs are known for their high and extremely stable light energy paired with very low energy consumption at no heat development and a long lifetime.
Intelligent LED Plugins: Adding new wavelengths (e.g. 340nm) or exchanging existing ones is as easy and safe. Auto-recognition of the individual plugins takes away the worry about filter positions in the reader.
Computer Controlled: LEDETECT 96 is fully computer controlled. Capture 96 (included) allows to read plates and collect the raw OD-data from the reader. With it's clipboard function, it allows to paste raw data into any spreadsheet program for further calculations.
MikroWin2000 compliant: In combination with MikroWin2000 data reduction software (optional), LEDETECT 96 adapts itself to any of your requirements for micropate based assays. Depending on your package selection, MikroWin2000 is the best choice for routine applications as well as extended screening, curvefit and kinetic studies.
| Plate Types: | 96 well |
| Optical system: | 8 channel Transmission Photometer |
| Light source: | Digital controlled LED lamps, wavelength specific |
| Photodetector: | 8 silicon Photodiodes |
| Wavelength range: | 340 - 750nm (special wavelengths up to 900nm) |
| Resolution: | 0.1 mOD (0.0001 OD) |
| Indication range: | 0.000 - 4.000 OD (Abs) |
| Accuracy: | ≤ ± 1% and ± 0.005 OD up to 2.5 OD (any wavelength) |
| Linearity: | ≤± 0.5% and ± 0.005 OD from 0.1 to 1.5 OD (any wavelength) ≤±0.75% from 1.5 to 2.5 OD (400nm – 750nm) ≤±0.75% and ± 0.005 OD from 0.1 to 2 OD (340 -400 nm) |
| Reproducibility: | ≤ ±0.3% at 1OD (any wavelength) ≤ ±0.5% at 2OD (400-750nm) |
| Measurement Mode: | Single and dual wavelength, Linear scan (30 points/well) for agglutination etc. |
| Reading speed: | 5 seconds (kinetic interval, single wavelength) 10 seconds (96 well, dual wavelength) |
| Wavelengths: | 4 wavelengths onboard (405, 450, 492, 620nm) up to 6 possible (340 – 750) |
| Shaking: | 4 speeds |
| Computer Interface: | USB 2.0 (USB 1.1 compatible) |
| PC Software: | Capture96 included MikroWin2000 trial version included |
| Dimensions / weight: | 23cm x 12cm x36cm (W x H x L), 6.7 kg net |
| Power Supply: | external power adapter 100-240V, 50 or 60 Hz (autosensing) 24VDC, 2.5A (approved to EN 60601-1-2, EN 61000-6-3, EN 61000-6-1,EN 60601-1, EN 60950) |
| Scope of Supply: | Power adapter, USB Cable, 4 Standard Filters, User manual (CD), Capture96 Software, MikroWin2000 (demo) |